Css Analytical Blog HP 5971 and HP 5972 Diffusion pump trouble shooting and service

If you have an HP 5971 or Agilent 5972 MSD and you can't get the diff pump to turn on there are only a few possibilities. 1) You have a massive leak and the diff pump controller board won't allow the diff pump to turn on. 2) The DC power supply in the MSD is dead and can't run the pump controller board, and 3) The diff pump heater is dead.